Sterilization Compatibility of Medical Polymers

The sterilization method is usually defined by what the device or instrument must do before a material is specified, and it can narrow the field of material options significantly. A polymer grade that looks like it will perform mechanically based on its technical data sheet properties is not a candidate if it cannot survive the sterilization cycle the device will see for its intended service life.

The Major Methods, and How They Can Affect Polymers

Steam Sterilization

Moist heat under pressure. The polymer has to tolerate a combination of high temperature, moisture and pressure in an autoclave. Devices that are designed to be reusable must stand up to repeated steam sterilization cycles. Hydrolytic stability at high temperatures over many cycles, not single cycle survival, is what qualifies material candidates for these types of instruments and devices. It is the requirement that drives specifications for polysulfone polymers including Udel PSU in lower temperature sterilization processes and Radel PPSU for reusable instrument trays and handles that go through higher temperature steam autoclaves. It is also the requirement that rules out less hydrolytically stable polymers such as polycarbonate.

Gamma irradiation

Penetrating, done after packaging, and attractive for that reason. The trade-off is that ionizing radiation can cause chain scission or crosslinking depending on the polymer, which may shift mechanical properties or color. Effects can be dose-dependent and can continue after exposure.

Ethylene oxide

Low temperature, which suits polymers that cannot take autoclave conditions. The considerations move from thermal and steam resistance to chemical tolerance: gas and residual permeation, aeration time, and residuals as a biological evaluation endpoint. This makes it a widely-used sterilization method for devices and instruments made from polymers like polycarbonate (PC) and nylon (PA) that are less tolerant to high temperature steam.

Electron beam

Also ionizing, with a shallower penetration profile than gamma at much shorter exposure times. Polymer response is broadly similar in kind to gamma irradiation.

How this affects material selection

Three questions settle most cases:

  • Single use or reusable? A single use device needs to survive one cycle, and many polymers can do this. A reusable instrument needs to survive its planned repeat-service life, and that requires materials that retain their performance over multiple sterilization cycles.
  • Is the method already fixed? If the device family is already validated on one sterilization method, the material has to fit the method rather than the other way around.
  • What changes first? Polymers rarely fail sterilization outright after a single or low number of cycles. They lose impact strength, discolor, or craze over time. Also, most polymers will exhibit more deterioration of a specific characteristic, such as color rather than toughness. Decide which of those is unacceptable relative to the device’s intended re-use life and number of sterilization cycles as step one to narrow material options.

Genesis grades by sterilization context

Sterilization and the biological evaluation

Sterilization is part of the device definition for biological evaluation under ISO 10993, because the finished, sterilized device is what gets evaluated. Changing the method late in the project forces a validation of the material’s tolerance to the sterilization process, and reopens part of the evaluation. See biocompatibility and ISO 10993.

Talk to our engineers before you fix the method

When the sterilization method and the material are both open decisions, it is the most efficient and least costly point in the specification process to get the pairing right. Contact our applications experts, or Request a Quote on the materials you’re considering.

FAQs: Sterilization Compatibility of Medical Polymers

PSU and PPSU polysulfones in particular are widely specified where repeated steam sterilization over a long re-use life is the governing requirement. Choosing the right sulfone polymer and grade depends on the intended life cycle of the device and its mechanical requirements and functions.

Polymer response to ionizing radiation varies by polymer grade and radiation dose. Confirm the specific grade, dose and acceptance criteria with Genesis engineering rather than generalizing from the polymer family.

It is possible but can be costly. The method is part of the device definition for biological evaluation, so changing it can reopen evaluation work as well as material qualification issues.

Genesis is a polymer conversion manufacturer. Sterilization is performed by the device manufacturer or its sterilization provider.